SimpleVIS®ViscometerReferenceGuideP82.0090
Emptying the Waste Bottle ReceiverCheck the volume of the Waste Bottle frequently to avoid spillage/overflow.Disconnect the Waste Bottle from the cap
either of the two following calibration intervals. Performance Based Method Determine need to calibrate by performing weekly check-runs of certified v
with the reference with the desired tolerance.Changing the temperature calibration by more than 0.05°C will affect the instrument viscosity calibratio
Unpack your SimpleVIS® ViscometerRemove all packing materials from the components. The SimpleVIS® Viscometer should be placed on a stable laboratory t
14vacuum line by approximately 2 inches.4. Connect the other ends of both pieces of tubing as indicated in the photo image.Select Target Temperature1
Waste BottleMedia CapP82.007515Printer Cable(optional)P82.0065Printer Power Supply(optional)Printer with Power Supply(optional)P82.0076PipettorP82.005
2139 High Tech Road • State College • PA • 16803814-353-8000 • 800-676-6232 • Fax [email protected] • www.cannoninstrument.comsi
Sample/Solvent PortRun Status Indicator LampsTouch Screen DisplayRS-232 SerialPort Vacuum LinePower ConnectorWaste LineExhaust MufflerDust CoverWaste
SAFETY CAUTIONS Please observe the following safety procedures and notices for proper operation of the SimpleVIS® Viscometer.• Make sure that you rea
The ~MAINS symbol indicates instructions or connections for the AC power supply. The AC Power input must match the electrical specifications listed
GETTING STARTEDPress the on switch on the AC/DC adapter. The Warm up Screen will display a “PLEASE WAIT” message. During this time, the instrument wil
Injecting a sample1. Using the supplied piston operated pipettor, insert the tip into a vial of test sample and pull the measured amount into the pip
Print the results (only if optional printer is installed)1. Press the Print button on the display screen to printout the test results to an attached s
Dry CyclePress the Dry button on the display wash screen to start the dry run cycle. The display will show the time remaining in the dry cycle.When th
The SimpleVIS® Viscometer was designed to require minimal maintenance through simplicity of design and built-in electronic diagnostics. All new instru