30
4-11. Parameters and default
Calibration with pure water
1: View Stored
Display measurement results — —
2: Set Sampling
Select sampling time 00 to 99s 10s
3: Select Oil
Table
Select temperature compensation table for oil
23A: Crude Oils 60/60°F
23B: Refined Products 60/60°F
23D: Lube Oils 60/60°F
53A : Crude Oils 15°C
53B : Refined Products 15°C
53D : Lube Oils 15°C
59A: Crude Oils 20°C
59B: Refined Products 20°C
23A/23B/23D/
53A/53B/53D/
59A/59B/59D
53A
4: Set Test
Set measurement temperature 15.0 to 25.0°C 20.0°C
5: Set Date &
Enter date and time. Date format: MM/DD/YY
Time format: HH:MM (24 hour format)
01/01/00 00:00 to
—
6: Set LCD
Adjust LCD brightness 0/25/50/75/100 50
7: Reset Test
Reset Sample Test No. to ‘01’ — 01
8: Clear Stored
Erase the memory of stored results. The D155 only
stores the last 10 measurement results.
— —
9: Export Stored
Data to USB
Export Stored Data from Memory to an MS-Excel
formatted file.
— —
10: Serial No. &
View D155 Serial Number and Firmware Version — —
CANNON
®
D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road • State College, PA • 16803 • USA
Comments to this Manuals